General information

The FIDAMAT 6 gas analyzer is suitable for determination of the total hydrocarbon content in air, in process gases, and high-boiling gas mixtures.

The SET FIDAMAT EX is used in the following areas:

  • Workplace monitoring
  • Purity checks and quality control
  • Safety monitoring
  • Quality assurance
  • Production of methanol, ethylene, propylene

The Set FIDAMAT EX is used in the following applications:

  • Chemicals and petrochemicals
  • Paint shops and dry-cleaning systems
  • Refineries
  • Solvent recovery systems

The FIDAMAT 6 gas analyzer is distinguished by its wide range of applications:

  • In the presence of up to 100 % H2O vapor
  • With high-boiling components (up to 200 °C)
  • In the presence of corrosive gases (with preliminary filter).

The FIDAMAT 6 exhibits:

  • Extremely low cross-sensitivity to interfering gases
  • Low consumption of combustion air
  • Low influence of oxygen on measured value

The analyzer is additionally equipped with warning and fault messages:

  • For failure of combustion gas
  • If the flame is extinguished
  • To indicate pump and filter faults

The Set FIDAMAT 6 EX additionally provides:

  • Use in hazardous areas of zone 2 in accordance with directive 94/EU
  • LEL measurement of explosive gas mixtures

Climatic conditions

 

Ambient temperature

 

  • Cabinet without air-conditioning

-20 … 30 °C

  • Cabinet with air-conditioning

-20 … 50 °C

  • Mounting plate

5 … 30 °C

Relative humidity

< 90%, non-condensing

Corrosive atmosphere

No

Auxiliary media

 

Compressed air

Approx. 6 000 hPa, purified instrument air, free of oil, water and dust

Volume flow

1 200 l/h

Power supply

 

Supply voltage

  • 100 … 120 V AC
  • 200 … 240 V AC

Frequency

48 … 63 Hz

Type of connections

 

Pipe material

Stainless steel

Connections

  • Metric (6 mm)
  • Imperial (1/4")

Cabling

 

Electrical design

According to IEC

Type of cables

Non-armored cables

Cable ID

No labeling of individual cores

Installation

 

Site

Interior

Ex-zone analyzer

ATEX II 3G

System design

 

Version

  • Basic unit in protective housing in accordance with ATEX
    610 x 600 x 630 mm (W x H x D)
  • Basic unit in protective housing on mounting plate, with processing of combustion air
    1 000 x 1 530 mm (W x H)
  • Basic unit on mounting plate in cabinet
    1 350 x 2 000 x 700 mm (W x H x D)

Degree of protection

IP54

Automatic calibration

Optional

Signal outputs

4 … 20 mA / floating contact

With sample gas return flow

No

Add-on electronics (option)

AUTOCAL extension

  • 8 additional binary inputs/outputs each
  • 8 additional binary inputs/outputs each and PROFIBUS PA

Measuring response

See technical specifications of FIDAMAT 6

Influencing variables

See technical specifications of FIDAMAT 6

Analyzer ID

Ex II 3G EEx n T3


FIDAMAT 6 without pump, with heated oven, with combustion air connection

Gases

Inlet pressure

Sample/calibration gas operating pressure

Flow through FID

Flow through bypass

 

 

Without

With

 

 

 

hPa (abs.)

hPa (abs.)

hPa

 

 

Combustion gas

3 000 ... 5 000

2 000 ± 20

~ 25

Combustion air

3 000 ... 5 000

1 485 ± 5

~ 320

~ 300

Sample gas

~ 2 000

1 500 ± 2

~ 3

~ 500

Zero gas

~ 2 000

1 500 ± 2

~ 3

~ 500

Calibration gas

~ 2 000

1 500 ± 2

~ 3

~ 500


Gas inlet conditions

The Set consists of the following components which can be selected as options:

  • FIDAMAT 6 in protective housing in accordance with ATEX (basic version)
  • FIDAMAT 6 in protective housing, mounted on mounting plate, and with combustion air preparation device
  • FIDAMAT 6 in protective housing, mounted on mounting plate, with combustion air preparation device, installed in a standard control cabinet

The basic version includes the FIDAMAT 6 in a purged sheet-steel protective housing for use in hazardous areas of Zone 2. The monitoring unit for monitoring the volume flow of the purging air and observation of the pressure limits in the protective housing is included in the delivery.

FIDAMAT 6 in protective housing

In a further version, the basic system can be ordered completely assembled on a mounting plate with dimensions 1 000 x 2 000 mm (W x H). This version additionally provides the complete combustion air preparation device, and is supplied with complete wiring and piping for installation in an analysis cabinet or standard protective cabinet.

FIDAMAT 6 in protective housing on mounting plate with combustion air preparation device

Legend

1

FIDAMAT 6

6

Gönnheimer EEx p monitoring

2

Combustion air preparation device

7

Terminal box

3

Pressure reducer for purging air

8

Reset switch

4

Pressure reducer for combustion air

9

Non-return valve

5

Pressure switch for combustion gas

10

Ball valves


The system can be optionally ordered completely ready for connection in a standard protective cabinet made of sheet-steel or FRP.

With all versions, the gas connections on the process side can be ordered with metric or Imperial dimensions.

Combustion gas supply

The combustion gas supply can be ordered optional to the various versions.

The components are fitted on a mounting plate, and include:

  • Pressure reducers required for combustion and calibration gases
  • Flow limiter for combustion gas
  • Solenoid valve for switching off the combustion gas in the event of a fault
  • Automatic switchover of combustion gas supply to replacement cylinder

The pressure of the combustion gas cylinder is monitored by a pressure switch. If the set limit is fallen below, automatic switchover to the replacement cylinder is carried out. An alarm is triggered at the same time to signal that it is necessary to replace the cylinder. The continuous supply with combustion gas means that interruption-free operation of the equipment is guaranteed.

Combustion and calibration gases do not belong to the scope of delivery. These must be ordered separately as required.

The sample gas is supplied to the FIDAMAT 6 at pressure, and passes via an anti-clogging fused silicone restrictor to the flame ionization detector.

Hydrogen (H2) is used as combustion gas. In the basic version, the FIDAMAT 6 is installed in an IP54 housing suitable for EEx(p) which is permanently purged with instrument air.

Purging prevents the formation of an explosive atmosphere in the housing should large leaks occur or if the combustion gas line breaks. To this end, any combustion gas released must be diluted down to a concentration < 25 % of the LEL. This requirement in accordance with DIN EN 60079-2 is achieved by permanent purging with instrument air at a flow rate of approx. 1 200 l/h.

The purging results in an overpressure in the protective housing. Therefore, any gas vapors which might occur occasionally cannot penetrate into the housing. The overpressure in the protective housing has a setpoint of 10 mbar. The monitored limits are between 0.8 and 15 mbar.

The purging flow rate for the protective housing and the set pressure limits are monitored.

In the event of a fault:

  • The power supply to the FIDAMAT 6 is switched off
  • The inputs/outputs of the FIDAMAT 6 are switched off by a Modex relay
  • The supply of combustion air is switched off by a solenoid valve outside the housing

The pressure of the combustion air supply for the FIDAMAT 6 is monitored by a pressure switch whose switching point is set to approx. 200 mbar below the supply pressure. If the pressure drops as a result of a leak or a break in the combustion gas line, the combustion air supply is switched off by a solenoid valve outside the housing.

Gases or gas mixtures that are frequently or continuously flammable must not be introduced into the FIDAMAT 6 gas analyzer.

The exhaust air and exhaust gas of the FIDAMAT 6 must be routed into the non-hazardous area.

Following installation and maintenance work affecting the gas paths, a leak test must be carried out as described in the manual.

FIDAMAT 6 in standard protective cabinet, dimensions in mm

  • Описание : set fidamat ex standard system for operation of an fidamat in a ex certified environment according to atex regulations