19" central unit производства Siemens

19" central unit

LDS 6, 19" central unit

Analytical performance

Measuring range

Adjustable

Detection limit (DL):

Calculated in accordance with VDI 2449, measured on every supplied analyzer during the temperature test (between 5 ... 45 °C) in accordance with VDI 4203.

Depending on sample gas component: see table for standard applications.

For Code ET and FT: in accordance with the requirements of 17th and 27th BImSchV

Smallest recommended measuring range (with 1 m path length)

Depending on sample gas component: see table for standard applications.

The maximum applicable measuring ranges can be found in the table of standard combinations. These can only be applied if the individual process conditions allow it. Please contact the Technical Support from Siemens for checking the applicability.

Accuracy

2% / 5%, depending on sample gas component and application letter. At best: detection limit. See table for standard applications.

For Code ET and FT: in accordance with the requirements of 17th and 27th BImSchV

Linearity

Better than 1%

Repeatability

2 % of the measured value or minimum detection limit (whichever is largest)

For Code ET and FT: in accordance with the requirements of 17th and 27th BImSchV

Calibration interval

No recalibration required thanks to internal reference cell

General information

Concentration units

ppmv, Vol%, mg/Nm3

Display

Digital concentration display (5 digits with floating decimal point)

Laser protection class

Class 1, safe to the eye

Certificates

CE marking, TÜV, MCERTS

Design, enclosure

Degree of protection

IP20 according to EN 60529

Dimensions

177 x 440 x 380 mm

Weight

Approx. 13 kg

Mounting

Horizontal

Electrical characteristics

Power supply

100 ... 240 V AC 50 ... 60 Hz, automatically adapted by the system; with a 3-channel central unit, an additional external power supply +24 V DC, 50 VA is included in the scope of delivery

Power consumption

50 W

EMC

According to EN 61326 and standard classification of NAMUR NE21

Electrical safety

According to EN 61010-1, overvoltage classification II

Fuse specifications

100 ... 240 V: T2.5L250V

Dynamic response

Warm-up time at 20 °C ambient temperature

Approx. 15 min

Response time

Less than 3 s, application-dependent

Integration time

1 … 100 s, adjustable

Influencing variables

Ambient temperature

< 0.5%/10 K of the measured value

Atmospheric pressure

Negligible

Process gas pressure compensation

Recommended

Process gas pressure range

See table for standard applications

Power supply changes

< 1%/30 V

Electrical inputs and outputs

Number of measurement channels

1 … 3, optional

Analog output

2 per channel, 4 ... 20 mA, floating, ohmic resistance max. 750 Ω

Analog inputs

2 per channel, designed for 4 ... 20 mA, 50 Ω

Binary outputs

6 per channel, with changeover contacts, configurable, 24 V AC/DC/1 A, floating

Binary inputs

6 per channel, designed for 24 V, floating, configurable

Communication interface

Ethernet 10BaseT (RJ-45)

Climatic conditions

Temperature range

5 … 45 °C during operation, -40 … +70 °C during storage and transportation

Atmospheric pressure

800 … 1 200 hPa

Humidity

< 85% relative humidity, above dew point (in operation and storage)


LDS 6, 19" central unit, dimensions in mm

Pin assignments

LDS 6, 19" central unit, pin assignments

Optical and electrical connections

LDS 6, three-channel 19" central unit, optical and electrical connections

The following table lists the measuring conditions for standard applications. The listed values for the measuring range and detection limit (DL) are only approximate values. The exact values at the respective measuring point depend on the totality of all influencing variables and can be determined by Siemens for the specific case. Please note that the values for the detection limit and the maximum measuring range refer to an optical path of 1 m. Longer path lengths will improve the detection limit, but not linearly. due to limiting effects such as dust load. The maximum applicable measuring ranges can only be used if permitted by the process conditions such as dust load.

Standard application

Effective optical path length: 0.3 … 12 m

Dust load3): < 50 g/Nm3

Process gas temperature

Tmin  … Tmax

Process gas pressure pmin  … pmax

Min. measuring range

(with 1 m eff. opt. path length)

Max. measuring range

(also dependent on eff. opt. path length: see following column)

(Max. measuring range x path length)

(DL x path length)

Accuracy4)

under standard conditions1) 2) without cross-interference of other gases

At 1 013 hPa with cross-interference of gas 2

Gas 1

Gas 2

Gas code

Appl. code

 

 

Gas 1

Gas 1

Gas 1

Gas 1

Gas 1

Gas 1

O2

 

A

B

600 … 1 200 °C

950 …1 050 hPa

0 … 15 vol%

0 … 100 vol%

240 vol%*m

0.3 vol%*m
At 600 °C

 

5 %

C

0 … 600 °C

950 …1 050 hPa

0 … 5 vol%

0 … 100 vol%

75 vol%*m

0.1 vol%*m

 

2 %5)

P

0 … 200 °C

950 …5 000 hPa

0 … 5 vol%

0 … 100 vol%

75 vol%*m

0.1 vol%*m

 

2 %

O2

Temp

B

B

600 … 1 200 °C

950 …1 050 hPa

0 … 35 vol%

0 … 100 vol%

240 vol%*m

0.7 vol%*m
At 600 °C

 

5 %

NH3

 

C

A

0 … 150 °C

950 …1 050 hPa

0 … 25 ppmv

0 … 500 ppmv

2 500 ppmv*m

0.5 ppmv*m

0.9 ppmv*m

At 15 vol% H2O, 55 °C

2 %

T

0 … 150 °C

950 …1 050 hPa

0 … 25 ppmv

0 … 500 ppmv

2 500 ppmv*m

0.5 ppmv*m

0.9 ppmv*m

At 15 vol% H2O, 55 °C

2 %

E

250 … 350 °C

950 …1 050 hPa

0 … 45 ppmv

0 … 500 ppmv

2 500 ppmv*m

0.9 ppmv*m

At 250 °C

1.4 ppmv*m

at 15 Vol% H2O, 250 °C

2 %

F

300 … 400 °C

950 …1 050 hPa

0 … 50 ppmv

0 … 500 ppmv

2 500 ppmv*m

1 ppmv*m

At 300 °C

1.5 ppmv*m

at 15 Vol% H2O, 300 °C

2 %

NH3

H2O

D

A

0 … 150 °C

950 …1 050 hPa

0 … 25 ppmv

0 … 100 ppmv

1 200 ppmv*m

0.5 ppmv*m

0.9 ppmv*m

At 15 vol% H2O, 55 °C

2 %

T

0 … 150 °C

950 …1 050 hPa

0 … 25 ppmv

0 … 100 ppmv

1 200 ppmv*m

0.5 ppmv*m

0.9 ppmv*m

At 15 vol% H2O, 55 °C

2 %

E

250 … 350 °C

950 …1 050 hPa

0 … 45 ppmv

0 … 100 ppmv

1 200 ppmv*m

0.9 ppmv*m

At 250 °C

1.4 ppmv*m

At 15 vol% H2O, 250 °C

2 %

F

300 … 400 °C

950 …1 050 hPa

0 … 50 ppmv

0 … 100 ppmv

1 200 ppmv*m

1 ppmv*m

At 300 °C

1.5 ppmv*m

At 15 vol% H2O, 300 °C

2 %

HCl

 

E

A

0 … 150 °C

950 …1 050 hPa

0 … 30 ppmv

0 … 6 000  ppmv

1 200 ppmv*m

0.6 ppmv*m

2.2 ppmv*m

At 15 % H2O, 55 °C

5 %

T

120 … 210 °C

950 …1 050 hPa

0 … 10 ppmv

0 … 60 ppmv

720 ppmv*m

 

 

 

H

150 … 250 °C

950 … 1 050 hPa

0 … 50 ppmv

0 … 6 000 ppmv

1 200 ppmv*m

1.0 ppmv*m

At 150 °C

3.1 ppmv*m

at 15 Vol% H2O, 150 °C

5 %

HCl

H2O

F

A

0 … 150 °C

950 …1 050 hPa

0 … 30 ppmv

0 … 100 ppmv

1 200 ppmv*m

0.6 ppmv*m

2.2 ppmv*m

At 15 % H2O, 55 °C

5 %

T

120 … 210 °C

950 …1 050 hPa

0 … 10 ppmv

0 … 60 ppmv

720 ppmv*m

 

 

 

H

150 … 250 °C

950 … 1 050 hPa

0 … 50 ppmv

0 … 100 ppmv

1 200 ppmv*m

1.0 ppmv*m

At 150 °C

3.1 ppmv*m

At 15 vol% H2O, 150 °C

5 %

HF

 

G

A

0 … 150 °C

950 …1 050 hPa

0 … 5 ppmv

0 … 1 500 ppmv

200 ppmv*m

0.1 ppmv*m

0.6 ppmv*m

At 15 vol% H2O, 55 °C

5 %

H

150 … 250 °C

950 … 1 050 hPa

0 … 5 ppmv

0 … 1 500 ppmv

200 ppmv*m

0.11 ppmv*m

At 150 °C

0.6 ppmv*m

At 15 vol% H2O, 150 °C

5 %

HF

H2O

H

A

0 … 150 °C

950 …1 050 hPa

0 … 5 ppmv

0 … 200 ppmv

200 ppmv*m

0.1 ppmv*m

0.6 ppmv*m

At 15 vol% H2O, 55 °C

5 %

H

150 … 250 °C

950 … 1 050 hPa

0 … 5 ppmv

0 … 200 ppmv

200 ppmv*m

0.11 ppmv*m

At 150 °C

0.6 ppmv*m

At 15 vol% H2O, 150 °C

5 %

CO

 

J

C

0 … 600 °C

950 …1 050 hPa

0 … 1.5 vol%

0 … 100 vol%

40 vol%*m

300 ppmv*m

1 500 ppmv*m

at 50 vol% CO2, 20 °C

2 %

CO

CO2

K

D

0 … 400 °C

800 …1 400 hPa

0 … 3.0 vol%

0 … 100 vol%

35 vol%*m

0.5 vol%*m

0.7 vol%*m

at 50 vol% CO2, 20 °C

2 %6)

CO2

 

L

A

0 … 150 °C

950 …1 050 hPa

0 … 7.5 vol%

0 … 100 vol%

40 vol%*m

300 ppmv*m

 

2 %

H2O

 

M

A

0 … 150 °C

950 …1 050 hPa

0 … 5 vol%

0 … 30 vol%

240 vol%*m

0.1 vol%*m

 

5 %

T

0 … 150 °C

950 …1 050 hPa

0 … 5 vol%

0 … 30 vol%

240 vol%*m

0.1 vol%*m

5 %


1) At 20 °C, 1 013 hPa

2) If the smallest permissible process gas temperature of the application is Tmin > 20 °C, the detection limit refers to Tmin and standard pressure (1 013 hPa)

3) At 0.3 m effective optical path length, average diameter of the dust particles: 15 µm, specific weight of the dust particles: 650 kg/m3

4) At least: Detection limit

5) Up to 200 °C, 5% above this

6) Up to 60 vol% CO or up to 60 vol% CO2. Higher CO or CO2 concentrations on request.

Standard application

Effective optical path length: 0.3 … 12 m

Dust load3): < 50 g/Nm3

Min. measuring range

(with 1 m eff. opt. path length)

Max. measuring range

(usually also dependent on eff. opt. path length: see following column)

(Max. measuring range x path length)

(DL x path length)

Accuracy5)

Purging gas mode

Purging gas medium

under standard conditions1) 2)

At 1 013 hPa with cross-interference of gas 1

Gas 1

Gas 2

Gas code

Appl. code

Gas 2

Gas 2

Gas 2

Gas 2

Gas 2

Gas 2

Standard

Optional

 

O2

 

A

B

 

 

 

 

 

 

E, F

G, H

Steam + air, N2

C

 

 

 

 

 

 

D

B

N2

P

 

 

 

 

 

 

D

B

N2

O2

Temp

B

B

 

600 … 1 200 °C

 

 

 

5)

F

H

Steam, N2

NH3

 

C

A


 

 

 

 

 

 

C

G

Air

T


 

 

 

 

 

 

C

G

Air

E


 

 

 

 

 

 

E

G

Air

F


 

 

 

 

 

 

E

G

Air

NH3

H2O

D

A


0 … 5 vol%

0 … 30 vol%

240 vol%*m

0.1 vol%*m

0.1 vol%*m

5 %

C

G

Air

T


0 … 5 vol%

0 … 30 vol%

240 vol%*m

0.1 vol%*m

0.1 vol%*m

5 %

C

G

Air

E


0 … 5 vol%

0 … 30 vol%

240 vol%*m

0.1 vol%*m

At 250 °C

0.1 vol%*m

At 250 °C

5 %

E

G

Air

F


0 … 5 vol%

0 … 30 vol%

240 vol%*m

0.1 vol%*m

at 300 °C"

0.1 vol%*m

At 300 °C

5 %

E

G

Air

HCl

 

E

A


 

 

 

 

 

 

C

G

Air

T


 

 

 

 

 

 

C

G

Air

H


 

 

 

 

 

 

E

G

Air

HCl

H2O

F

A


0 … 5 vol%

0 … 30 vol%

360 vol%*m

0.1 vol%*m

0.1 vol%*m

5 %

C

G

Air

T


0 … 5 vol%

0 … 30 vol%

360 vol%*m

 

 

 

C

G

Air

H


0 … 5 vol%

0 … 30 vol%

360 vol%*m

0.1 vol%*m

At 150 °C

0.1 vol%*m

At 150 °C

5 %

E

G

Air

HF

 

G

A


 

 

 

 

 

 

C

G

Air

H


 

 

 

 

 

 

E

G

Air

HF

H2O

H

A


0 … 5 vol%

0 … 30 vol%

360 vol%*m

0.1 vol%*m

0.1 vol%*m

5 %

C

G

Air

H


0 … 5 vol%

0 … 30 vol%

360 vol%*m

300 ppmv*m

At 200 °C

300 ppmv*m

At 200 °C

5 %

E

G

Air

CO

 

J

C


 

 

 

 

 

 

E

G

Air, N2

CO

CO2

K

D

0 … 7.5 vol%

0 … 100 vol%

75 vol%*m

0.5 vol%*m

1 500 ppmv*m

At 50 vol% CO, 20 °C

2 … 5 %6)

C

G

Air

CO2

 

L

A

 

 

 

 

 

 

C

G

Air

H2O

 

M

A

 

 

 

 

 

 

C

G

Air

T

 

 

 

 

 

 

C

G

Air


1) At 20 °C, 1 013 hPa

2) If the smallest permissible process gas temperature of the application is Tmin > 20 °C, the detection limit refers to Tmin and standard pressure (1 013 hPa)

3) At 0.3 m optical path length, average diameter of the dust particles: 15 µm, specific weight of the dust particles: 650 kg/m3

4) At least: Detection limit

5) At 600 °C and at least 5 vol%*m O2 concentration: Resolution = 15 °C, at 1 000 °C and at least 5 vol%*m O2 concentration: Resolution = 25 °C

6) Depends on temperature (higher values at higher temperatures)

Код Заказной номер Описание Вес (кг) Заказать
70854 7MB6121-.....-.... lds 6 диоднолазерный спектрометр, центральный модуль 19 Заказать
78125 A5E00534675 calibration kit nh3, version 1 3.6 Заказать
124603 A5E00823339003 calibration kit rc3009 tuev/ mcert, nh3, 2 cells, version 1 service calibration set for analysers certified acc. to 17 th bimschv. this set is suitable to perform regular linearity checks as required from the test authority 4.8 Заказать
78145 A5E00823339004 calibration kit rc3009,tuev/ mcert, nh3/h2o, 3 cells, version 1, service calibration set for analysers certified acc. to 17 th bimschv. this set is suitable to perform regular linearity checks as required from the test authority 5.8 Заказать
124604 A5E00823339005 calibration kit rc3009 tuev/ mcert, h2o, 2 cells version 1 service calibration set for analysers certified acc. to 17 th bimschv. this set is suitable to perform regular linearity checks as required from the test authority 4.8 Заказать
124605 A5E00823339008 tuev/mcert linearity check kit hcl, 2 cells 4.7 Заказать
78148 A5E00823339009 tuev/mcert linearity check kit hcl/h2o, 3 cells 5.8 Заказать
78150 A5E00823339013 tuev/mcert linearity check kit nh3, version 2, 2 cells 4.7 Заказать
124607 A5E00823339014 tuev/mcert linearity check kit h2o/nh3, version 2, 3 cells 5.8 Заказать
78151 A5E00823339015 tuev/mcert linearity check kit h2o, version 2, 2 cells 4.7 Заказать
78172 A5E00854188 lds6, 3 channel external power supply 3 Заказать
124664 A5E01075594 calibration verification kit nh3, version 2 3.6 Заказать
;