Field device производства Siemens

Field device

General (based on DIN EN 61207/IEC 1207. All data refers to the binary gas mixture H2 in N2)

Measuring ranges

4, internally and externally switchable; automatic measuring range switchover also possible

Span

Application-dependent (see ordering data)

Measuring ranges with suppressed zero point

Application-dependent (see ordering data)

Operating position

Front wall, vertical

Conformity

CE marking in accordance with EN 50081-1/EN 50081-2 and RoHS

Design, enclosure

Degree of protection

IP65 according to EN 60529

Weight

Approx. 25 kg

Electrical characteristics

EMC (Electromagnetic Compatibility)

In accordance with standard requirements of NAMUR NE21 (08/98) and EN 61326

Electrical safety

In accordance with EN 61010-1; overvoltage category II

Power supply (see nameplate)

100 AC -10 % ... 120 V AC +10 %, 47 ... 63 Hz

or

200 AC -10 % ... 240 V AC +10 %, 47 ... 63 Hz

Power consumption

  • Approx. 25 VA (gas connection block unheated)
  • Approx. 330 VA (gas connection block heated)

Fuse values (gas connection unheated)

100 ... 120 V

F3 1T/250

F4 1T/250

200 ... 240 V

F3 0.63T/250

F4 0.63T/250

Fuse values (gas connection heated)

100 ... 120 V

F1 1T/250

F2 4T/250

F3 4T/250

F4 4T/250

200 ... 240 V

F1 0.63T/250

F2 2.5T/250

F3 2.5T/250

F4 2.5T/250

Gas inlet conditions

 

Sample gas pressure

800 ... 1 100 hPa (absolute)

Sample gas flow

30 ... 90 l/h

Sample gas temperature

Min. 0 to max. 50 °C, but above the dew point

Temperature

 

  • of the measuring cell (sensor)

70 ºC

  • of the measureming cell block (base)

80 ºC (heated)

Sample gas humidity

< 90% relative humidity

Purging gas pressure

 

  • Permanent

165 hPa above ambient pressure

  • For short periods

Max. 250 hPa above ambient pressure

Dynamic response(the dynamic and measuring response refers to the measurement of H2 in N2)
(referred to sample gas pressure 1 000 hPa absolute, sample gas flow 0.5 l/min, and ambient temperature 25 °C)

Warm-up period

< 30 min at room temperature (the technical specification will be met after 2 hours)

Delayed display (T90)

Approx. 35 s (including dead time)

Electrical damping

0 ... 100 s, parameterizable

Dead time (the diffusion to the probes is the determining variable)

Approx. 34 s

Measuring response(the dynamic and measuring response refers to the measurement of H2 in N2)
(referred to sample gas pressure 1 000 hPa absolute, sample gas flow 0.5 l/min, and ambient temperature 25 °C)

Output signal fluctuation (3σ value)

< ± 1% of the smallest possible span according to rating plate with electronic damping constant of 1 s

Zero point drift

< ± 1% of the current span/week

Measured-value drift

< ± 1% of the smallest possible span (according to rating plate)/week

Repeatability

< ± 1% of the current span

Detection limit

1% of the smallest possible span according to rating plate

Linearity error

< ± 1% of the current span

Influencing variables (referred to sample gas pressure 1 000 hPa absolute, sample gas flow 0.5 l/min, and ambient temperature 25 °C)

 

Ambient temperature

< 2%/10 K referred to smallest possible span according to rating plate

Accompanying gases

Deviation from zero point (for influence of interfering gas, see section "Cross-interference")

Sample gas flow

0.2% of the current measuring span with a change in flow of 0.1 l/min within the permissible flow range

Sample gas pressure

< 1 % of the span with a change in pressure of 100 hPa

Power supply

< 0.1% of the output signal span with rated voltage ± 10%

Electrical inputs and outputs

 

Analog output

0/2/4 ... 20 mA, isolated; max. load 750 Ω

Relay outputs

6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated

Analog inputs

2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of cross-interference

Binary inputs

6, designed for 24 V, isolated, freely parameterizable, e.g. for measuringrange switchover

Serial interface

RS 485

Options

AUTOCAL function with 8 additional binary inputs and 8 additional relay outputs, also with PROFIBUS PA (on request) or PROFIBUS DP (on request)

Climatic conditions

 

Permissible ambient temperature

-40 ... +70 °C during storage and transportation, 5 … 45 °C during operation

Permissible humidity (dew point must not be fallen below)

< 90 % relative humidity as annual average, during storage and transportation


CALOMAT 62, field device, dimensions in mm

Pin assignment (electrical and gas connections)

CALOMAT 62, field device, pin and terminal assignment

CALOMAT 62, field device, pin and terminal assignment of the AUTOCAL board and PROFIBUS connectors

CALOMAT 62, field device, gas connections and electrical connections

Код Заказной номер Описание Вес (кг) Заказать
70850 7MB2531-.....-.... calomat 62 анализатор для полевого монтажа 30 Заказать
77978 A5E00057315 additional electronics profibus-pa (for field units) 0.5 Заказать
124414 A5E00057318 addtional electronics with 8 binary inputs and 8 binary outputs and profibus dp 0.5 Заказать
124417 A5E00064223 additional electronics with 8 binary inputs and 8 binary outputs (retrofit set, series 6, field version) 0.35 Заказать
78169 A5E00852382 usb-rs485 converter 1 Заказать
124619 A5E00852383 rs 485 / ethernet-converter 1 Заказать
126081 C79451-Z1589-U1 interface converter rs 485 / rs 232 (retrofit set) 0.716 Заказать
;