General (based on DIN EN 61207/IEC 1207. All data refers to the binary gas mixture H2 in N2) | |
Measuring ranges | 4, internally and externally switchable; automatic measuring range switchover also possible |
Span | Application-dependent (see ordering data) |
Measuring ranges with suppressed zero point | Application-dependent (see ordering data) |
Operating position | Front wall, vertical |
Conformity | CE marking in accordance with EN 50081-1/EN 50081-2 and RoHS |
Design, enclosure | |
Degree of protection | IP65 according to EN 60529 |
Weight | Approx. 25 kg |
Electrical characteristics | |
EMC (Electromagnetic Compatibility) | In accordance with standard requirements of NAMUR NE21 (08/98) and EN 61326 |
Electrical safety | In accordance with EN 61010-1; overvoltage category II |
Power supply (see nameplate) | 100 AC -10 % ... 120 V AC +10 %, 47 ... 63 Hz or 200 AC -10 % ... 240 V AC +10 %, 47 ... 63 Hz |
Power consumption |
|
Fuse values (gas connection unheated) | 100 ... 120 V F3 1T/250 F4 1T/250 200 ... 240 V F3 0.63T/250 F4 0.63T/250 |
Fuse values (gas connection heated) | 100 ... 120 V F1 1T/250 F2 4T/250 F3 4T/250 F4 4T/250 200 ... 240 V F1 0.63T/250 F2 2.5T/250 F3 2.5T/250 F4 2.5T/250 |
Gas inlet conditions | |
Sample gas pressure | 800 ... 1 100 hPa (absolute) |
Sample gas flow | 30 ... 90 l/h |
Sample gas temperature | Min. 0 to max. 50 °C, but above the dew point |
Temperature | |
| 70 ºC |
| 80 ºC (heated) |
Sample gas humidity | < 90% relative humidity |
Purging gas pressure | |
| 165 hPa above ambient pressure |
| Max. 250 hPa above ambient pressure |
Dynamic response(the dynamic and measuring response refers to the measurement of H2 in N2) | |
Warm-up period | < 30 min at room temperature (the technical specification will be met after 2 hours) |
Delayed display (T90) | Approx. 35 s (including dead time) |
Electrical damping | 0 ... 100 s, parameterizable |
Dead time (the diffusion to the probes is the determining variable) | Approx. 34 s |
Measuring response(the dynamic and measuring response refers to the measurement of H2 in N2) | |
Output signal fluctuation (3σ value) | < ± 1% of the smallest possible span according to rating plate with electronic damping constant of 1 s |
Zero point drift | < ± 1% of the current span/week |
Measured-value drift | < ± 1% of the smallest possible span (according to rating plate)/week |
Repeatability | < ± 1% of the current span |
Detection limit | 1% of the smallest possible span according to rating plate |
Linearity error | < ± 1% of the current span |
Influencing variables (referred to sample gas pressure 1 000 hPa absolute, sample gas flow 0.5 l/min, and ambient temperature 25 °C) | |
Ambient temperature | < 2%/10 K referred to smallest possible span according to rating plate |
Accompanying gases | Deviation from zero point (for influence of interfering gas, see section "Cross-interference") |
Sample gas flow | 0.2% of the current measuring span with a change in flow of 0.1 l/min within the permissible flow range |
Sample gas pressure | < 1 % of the span with a change in pressure of 100 hPa |
Power supply | < 0.1% of the output signal span with rated voltage ± 10% |
Electrical inputs and outputs | |
Analog output | 0/2/4 ... 20 mA, isolated; max. load 750 Ω |
Relay outputs | 6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated |
Analog inputs | 2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of cross-interference |
Binary inputs | 6, designed for 24 V, isolated, freely parameterizable, e.g. for measuringrange switchover |
Serial interface | RS 485 |
Options | AUTOCAL function with 8 additional binary inputs and 8 additional relay outputs, also with PROFIBUS PA (on request) or PROFIBUS DP (on request) |
Climatic conditions | |
Permissible ambient temperature | -40 ... +70 °C during storage and transportation, 5 … 45 °C during operation |
Permissible humidity (dew point must not be fallen below) | < 90 % relative humidity as annual average, during storage and transportation |
CALOMAT 62, field device, dimensions in mm
CALOMAT 62, field device, pin and terminal assignment
CALOMAT 62, field device, pin and terminal assignment of the AUTOCAL board and PROFIBUS connectors
CALOMAT 62, field device, gas connections and electrical connections
Код | Заказной номер | Описание | Вес (кг) | Заказать |
---|---|---|---|---|
70850 | 7MB2531-.....-.... | calomat 62 анализатор для полевого монтажа | 30 | Заказать |
77978 | A5E00057315 | additional electronics profibus-pa (for field units) | 0.5 | Заказать |
124414 | A5E00057318 | addtional electronics with 8 binary inputs and 8 binary outputs and profibus dp | 0.5 | Заказать |
124417 | A5E00064223 | additional electronics with 8 binary inputs and 8 binary outputs (retrofit set, series 6, field version) | 0.35 | Заказать |
78169 | A5E00852382 | usb-rs485 converter | 1 | Заказать |
124619 | A5E00852383 | rs 485 / ethernet-converter | 1 | Заказать |
126081 | C79451-Z1589-U1 | interface converter rs 485 / rs 232 (retrofit set) | 0.716 | Заказать |