;

19" rack unit производства Siemens

19" rack unit

ULTRAMAT/OXYMAT 6, 19" rack unit

General information

 

Operating position

Front wall, vertical

Conformity

CE mark in accordance with EN 50081-1, EN 50082-2

Design, enclosure

Weight

Approx. 21 kg

Degree of protection

IP20 according to EN 60529

Electrical characteristics

EMC (Electromagnetic Compatibility)

In accordance with standard requirements of NAMUR NE21 (08/98)

Electrical safety

According to EN 61010-1, overvoltage category III

Power supply

100 ... 120 V AC (nominal range of use 90 ... 132 V), 48 ... 63 Hz or
200 ... 240 V AC (nominal range of use 180 ... 264 V), 48 ... 63 Hz

Power consumption

Approx. 70 VA

Fuse values

120 ... 120 V: F1/F2 = T 1.6 A
200 ... 240 V: F1/F2 = T 1 A

Electrical inputs and outputs (per channel)

Analog output

0/2/4 ... 20 mA, isolated; max. load 750 Ω

Relay outputs

6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated, non-sparking

Analog inputs

2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of influence of accompanying gas (correction of cross-interference)

Binary inputs

6, designed for 24 V, isolated, freely parameterizable, e.g. for measuring range switchover

Serial interface

RS 485

Options

AUTOCAL function each with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP

Climatic conditions

 

Permissible ambient temperature

-30 ... +70 °C during storage and transportation, 5 … 45 °C during operation

Permissible humidity

< 90% relative humidity, during storage and transportation (dew point must not be undershot)


Technical data, ULTRAMAT channel

Measuring ranges

4, internally and externally switchable; autoranging is also possible

Smallest possible measuring range

Dependent on the application, e.g.
CO: 0 ... 10 vpm
CO2: 0 ... 5 vpm

Largest possible measuring range

Dependent on the application

Measuring ranges with suppressed zero point

Any zero point within 0 ... 100 vol.% can be implemented; smallest possible span 20 %

Characteristic

Linearized

Influence of interfering gases must be considered separately

Gas inlet conditions

Permissible sample gas pressure

 

  • Without pressure switch

700 ... 1 500 hPa (absolute)

  • With integrated pressure switch

700 ... 1 300 hPa (absolute)

Sample gas flow

18 ... 90 l/h (0.3 ... 1.5 l/min)

Sample gas temperature

Min. 0 to max. 50 °C, but above the dew point

Sample gas humidity

< 90% (relative humidity), or dependent on measuring task, non-condensing

Dynamic response

Warm-up period

At room temperature < 30 min (the technical specification will be met after 2 hours)

Delayed display (T90-time)

Dependent on length of analyzer chamber, sample gas line and parameterizable damping

Damping (electrical time constant)

0 ... 100 s, parameterizable

Dead time (purging time of the gas path in the unit at 1 l/min)

Approx. 0.5 ... 5 s, depending on version

Time for device-internal signal processing

< 1 s

Pressure correction range

 

Pressure sensor

 

  • Internal

700 ... 1 200 hPa absolute

  • External

700 ... 1 500 hPa absolute

Measuring response (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< ± 1% of the smallest possible measuring range according to rating plate

Zero point drift

<± 1% of the current measuring range/week

Measured-value drift

<± 1% of the current measuring range/week

Repeatability

≤ 1% of the current measuring range

Detection limit

1 % of the smallest possible measuring range

Linearity error

< 0.5% of the full-scale value

Influencing variables (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Ambient temperature

< 1% of current measuring range/10 K (with constant receiver cell temperature)

Sample gas pressure

  • When pressure compensation has been switched on: < 0.15% of the span/1% change in atmospheric pressure
  • When pressure compensation has been switched off: < 1.5% of the span/1% change in atmospheric pressure

Sample gas flow

Negligible

Power supply

< 0.1% of the current measuring range with rated voltage ± 10%

Environmental conditions

Application-specific measuring influences possible if ambient air contains measured component or cross interference-sensitive gases


Technical data, OXYMAT channel

Measuring ranges

4, internally and externally switchable; automatic measuring range switchover also possible

Smallest possible span (relating to sample gas pressure 1 000 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

0.5 vol.%, 2 vol.% or 5 vol.% O2

Largest possible measuring range

100 vol. % O2

Measuring ranges with suppressed zero point

Any zero point within 0 ... 100 vol.% can be implemented, provided that a suitable reference gas is used

Gas inlet conditions

Permissible sample gas pressure

 

  • With pipes

500 ... 3 000 hPa absolute

  • With hoses
 

  • Without pressure switch

500 ... 1 500 hPa absolute

  • With pressure switch

500 ... 1 300 hPa absolute

Sample gas flow

18 ... 60 l/h (0.3 ... 1 l/min)

Sample gas temperature

0 ... 50 ºC

Sample gas humidity

< 90% RH (relative humidity)

Reference gas pressure (high-pressure version)

2 000 ... 4 000 hPa above sample gas pressure, but max. 5 000 hPa

Reference gas pressure (low-pressure version)

Min. 100 hPa above sample gas pressure

Dynamic response

Warm-up period

At room temperature < 30 min (the technical specification will be met after 2 hours)

Delayed display (T90 time)

Min. 1.5 ... 3.5 s, depending on version

Damping (electrical time constant)

0 ... 100 s, parameterizable

Dead time (purging time of the gas path in the unit at 1 l/min)

Approx. 0.5 ... 2.5 s, depending on version

Time for device-internal signal processing

< 1 s

Pressure correction range

Pressure sensor

 

  • Internal

500 ... 2 000 hPa absolute

  • External

500 ... 3 000 hPa absolute

Measuring response (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (corresponds to ± 0.25% at 2σ)

Zero point drift

< 0.5%/month of the smallest possible measuring span according to rating plate

Measured-value drift

≤ 0.5%/month of the current measuring range

Repeatability

≤ 1%/month of the current measuring range

Detection limit

1% of the current measuring range

Linearity error

1% of the current measuring range

Influencing variables (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Ambient temperature

  • < 0.5%/10 K referred to smallest possible span according to rating plate
  • With measuring span 0.5%: 1%/10 K

Sample gas pressure (with air (100 hPa) as reference gas, correction of the atmospheric pressure fluctuations is only possible if the sample gas can vent to ambient air)

  • When pressure compensation has been switched off: < 2% of the current measuring range/1% atmospheric pressure change
  • When pressure compensation has been switched on: < 0.2% of the current measuring range/1% atmospheric pressure change

Accompanying gases

Deviation from zero point corresponding to paramagnetic or diamagnetic deviation of accompanying gas

Sample gas flow

< 1% of the smallest possible span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range

Power supply

< 0.1 % of the current measuring range with rated voltage ± 10%


ULTRAMAT/OXYMAT 6, 19“ unit, dimensions in mm

Pin assignment (electrical and gas connections)

ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment

ULTRAMAT/OXYMAT 6, 19“ unit, pin assignment of AUTOCAL board and PROFIBUS connectors

ULTRAMAT/OXYMAT 6, 19“ unit, gas and electrical connections

Код Заказной номер Описание Вес (кг) Заказать
117339 7MB2023-.....-.... анализатор ultramat/oxymat 6 исполнение 19" для монтажа в шкаф 21 Заказать
70843 7MB2024-.....-.... анализатор ultramat 6 для измерения компонентов в инфрокрасном диапазоне, шкафное исполнение 21 Заказать
77977 A5E00057307 autocal board with 8 binary- inputs/outputs and profibus pa 0.5 Заказать
124413 A5E00057312 autocal board with mit 8 binary inputs and 8 binary outputs and profibus dp 0.5 Заказать
78169 A5E00852382 usb-rs485 converter 1 Заказать
124619 A5E00852383 rs 485 / ethernet-converter 1 Заказать
125988 C79000-G5200-C143 manual ultramat 6 / oxymat 6 0.42 Заказать
79601 C79000-G5272-C143 manual (italian) ultramat 6e / oxymat 6e 0.42 Заказать
79602 C79000-G5276-C143 manual (english) ultramat 6 / oxymat 6 0.8 Заказать
79604 C79000-G5277-C143 manual (french) ultramat6 / oxymat6 0.335 Заказать
79606 C79000-G5278-C143 manual (spanish) ultramat6 / oxymat6 0.41 Заказать
79692 C79451-A3480-D33 retrofit kit serial interface for automotiv (ak)for ultramat 6e 2p and ultramat/oxymat 6e 0.25 Заказать
126071 C79451-A3480-D511 additional electronics modul 0.243 Заказать
126081 C79451-Z1589-U1 interface converter rs 485 / rs 232 (retrofit set) 0.716 Заказать
;