19" rack unit

ULTRAMAT 6, 19" rack unit

 

General information

 

Measuring ranges

4, internally and externally switchable; autoranging is also possible

Smallest possible measuring range

Dependent on the application: e.g. CO: 0 … 10 vpm, CO2: 0 … 5 vpm

Largest possible measuring span

Dependent on the application

Measuring range with suppressed zero point

Any zero point within 0 … 100 vol.% can be implemented; smallest possible span 20%

Operating position

Front wall, vertical

Conformity

CE mark in accordance with EN 50081-1, EN 50082-2

Influence of interfering gases must be considered separately

Design, enclosure

 

Weight

Approx. 15 kg (with one IR channel)

Approx. 21 kg (with two IR channels)

Degree of protection

IP20 according to EN 60529

Electrical characteristics

 

EMC (Electromagnetic Compatibility)

In accordance with standard requirements of NAMUR NE21 (08/98)

Electrical safety

According to EN 61010-1, overvoltage category III

Power supply

100 … 120 V AC (nominal range of use 90 … 132 V), 47 … 63 Hz or

200 … 240 V AC (nominal range of use 180 … 264 V), 47 … 63 Hz

Power consumption

1-channel unit: Approx. 40 VA

2-channel unit: Approx. 70 VA

Fuse values

 

  • 100 ... 120 V

1 T/250 (7MB2121)
1.6 T/250 (7MB2123)

  • 200 ... 240 V

0.63 T/250 (7MB2121)
1 T/250 (7MB2123)

Gas inlet conditions

 

Permissible sample gas pressure

 

  • With hoses
 

  • Without pressure switch

600 … 1 500 hPa (absolute)

  • With pressure switch

700 … 1 300 hPa (absolute)

  • With pipes (without pressure switch)

600 … 1 500 hPa (absolute)

Sample gas flow

18 … 90 l/h (0.3 … 1.5 l/min)

Sample gas temperature

Min. 0 ... max. 50 °C, but above the dew point

Sample gas humidity

< 90% RH (relative humidity), or dependent on measuring task, non-condensing

Dynamic response

 

Warm-up period

At room temperature < 30 min (the technical specification will be met after 2 hours)

Delayed display (T90-time)

Dependent on length of analyzer chamber, sample gas line and parameterizable damping

Damping (electrical time constant)

0 … 100 s, parameterizable

Dead time (purging time of the gas path in the unit at 1 l/min)

Approximately 0.5 ... 5 s, depending on version

Time for device-internal signal processing

< 1 s

Pressure correction range

 

Pressure sensor

 

  • Internal

700 … 1 200 hPa absolute

  • External

700 … 1 500 hPa absolute

Measuring response (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< ± 1% of the smallest possible measuring range according to rating plate

Zero point drift

<± 1% of the current measuring range/week

Measured-value drift

<± 1% of the current measuring range/week

Repeatability

≤ 1% of the current measuring range

Detection limit

1 % of the smallest possible measuring range

Linearity error

< 0.5% of the full-scale value

Influencing variables (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Ambient temperature

< 1% of current measuring range/10 K (with constant receiver cell temperature)

Sample gas pressure

  • When pressure compensation has been switched on: < 0.15% of the span/1% change in atmospheric pressure
  • When pressure compensation has been switched off: < 1.5% of the span/1% change in atmospheric pressure

Sample gas flow

Negligible

Power supply

< 0.1% of the current measuring range with rated voltage ± 10%

Environmental conditions

Application-specific measuring influences possible if ambient air contains measured components or cross interference-sensitive gases

Electrical inputs and outputs

 

Analog output

0/2/4 ... 20 mA, isolated; load ≤ 750 Ω

Relay outputs

6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated, non-sparking

Analog inputs

2, dimensioned for 0/2/4 ... 20 mA for external pressure sensor and accompanying gas influence correction (correction of cross-interference)

Binary inputs

6, designed for 24 V, isolated, freely parameterizable, e.g. for measuring range switchover

Serial interface

RS 485

Options

AUTOCAL function with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP

Climatic conditions

 

Permissible ambient temperature

-30 ... +70 °C during storage and transportation,
5 ... 45 °C during operation

Permissible humidity

< 90% RH (relative humidity) as annual average, during storage and transportation (dew point must not be undershot)


ULTRAMAT 6, 19“ unit, dimensions in mm (example: 1-channel version)

Pin assignment (electrical and gas connections)

ULTRAMAT 6, 19“ unit, pin assignment

ULTRAMAT 6, 19“ unit, pin assignment of AUTOCAL board and PROFIBUS connectors

ULTRAMAT 6, 19“ unit, gas and electrical connections (example: 2-channel version)

Код Заказной номер Описание Вес (кг) Заказать
70845 7MB2121-.....-.... ultramat 6e газовый анализатор одноканальный, 19" для установки в шкаф 15 Заказать
117343 7MB2123-.....-.... ultramat 6 газовый анализатор двухканальный, 19" для установки в шкаф 21 Заказать
70846 7MB2124-.....-.... ultramat 6 газовый анализатор, 19" для установки в шкаф, для измерения 2-3 газов 30 Заказать
77977 A5E00057307 autocal board with 8 binary- inputs/outputs and profibus pa 0.5 Заказать
124413 A5E00057312 autocal board with mit 8 binary inputs and 8 binary outputs and profibus dp 0.5 Заказать
78169 A5E00852382 usb-rs485 converter 1 Заказать
124619 A5E00852383 rs 485 / ethernet-converter 1 Заказать
125988 C79000-G5200-C143 manual ultramat 6 / oxymat 6 0.42 Заказать
79601 C79000-G5272-C143 manual (italian) ultramat 6e / oxymat 6e 0.42 Заказать
79602 C79000-G5276-C143 manual (english) ultramat 6 / oxymat 6 0.8 Заказать
79604 C79000-G5277-C143 manual (french) ultramat6 / oxymat6 0.335 Заказать
79606 C79000-G5278-C143 manual (spanish) ultramat6 / oxymat6 0.41 Заказать
126029 C79165-A3044-B176 temperature sensor 0.1 Заказать
79692 C79451-A3480-D33 retrofit kit serial interface for automotiv (ak)for ultramat 6e 2p and ultramat/oxymat 6e 0.25 Заказать
126071 C79451-A3480-D511 additional electronics modul 0.243 Заказать
126072 C79451-A3480-D512 additional electronics for the automotive industrie (ak) 0.099 Заказать
126081 C79451-Z1589-U1 interface converter rs 485 / rs 232 (retrofit set) 0.716 Заказать
;